50 Ω Termination Resistor for EIS Signal Generator

Installation

Applies to EIS systems using a Gantner Instruments signal generator

At the rear of the EIS Q.core, it is recommended to install a 50 Ω terminator (using a BNC T-connector) for the signal generator connection, as shown below.

eis_signal_generator_50ohm_bnc_terminator_1_edit

EIS plugin setting

Go to SystemSetup → Signal Generator. In the Load-RefGen field, enter 50. The amplitude will then be scaled accordingly.

 

260505_eis_50Ohm_terminator_2511199

The setting is available in EIS Engine version 1.13.0 and later.

Article numbers