Skip to content
English
  • There are no suggestions because the search field is empty.

How to set up order tracking in GI.bench

Measure vibration against shaft speed, configure a Q.station FFT evaluator, and visualise the results in a GI.bench dashboard.

Order tracking shows which vibration components follow shaft speed and how their amplitudes change during a run-up or coast-down. With Gantner’s Q.series X system, acquire vibration and a shaft-speed reference together. Use the Order Tracking Heatmap in GI.bench to see the speed-dependent spectrum, or calculate selected order results on a compatible Q.station with its FFT Processor.

What is an order?

An order is a frequency expressed as a multiple of a reference shaft’s rotational frequency. The first order (1×) occurs once per revolution; 2× occurs twice. Convert shaft speed to frequency using shaft frequency [Hz] = speed [rpm] ÷ 60. A component’s order equals its frequency divided by the shaft frequency.

At 1,800 rpm the 1× and 2× components are at 30 and 60 Hz. At 3,600 rpm they move to 60 and 120 Hz, but their orders stay the same. A fixed-frequency resonance, by contrast, stays near the same frequency as speed changes. An order crossing that frequency may excite it, producing a local increase in amplitude.

Use order tracking to compare run-ups and coast-downs, identify the speed range in which a vibration becomes pronounced, and distinguish rotation-related components from fixed-frequency effects. A strong 1× component can be consistent with imbalance, but it does not identify a fault on its own. For systems with several shafts, document which shaft defines 1×.

Worked 1× example: At 480 rpm, check for the shaft-related peak near 8 Hz. At 840 rpm, look for the same 1× component near 14 Hz. For each FFT, the evaluator writes the amplitude and the detected frequency to their assigned channels. A trend of the amplitude channel shows how the first-order vibration changes during the run; the frequency channel lets you check that the tracked peak follows shaft speed.

1. Acquire vibration and shaft speed

  1. Configure the vibration channel. Connect an accelerometer to a compatible input, with the correct sensor supply, range, scaling and units. Use a sample rate and anti-alias filtering appropriate for the highest frequency of interest. For example, 10× at 6,000 rpm is 1,000 Hz: the theoretical minimum sample rate is above 2,000 samples/s, and the practical rate must also allow for the input filter and usable bandwidth.
  2. Configure the speed reference. Use a tachometer, encoder or other suitable speed signal. If the input measures pulse frequency, divide it by the number of pulses counted per shaft revolution to obtain mechanical frequency in Hz. Multiply by 60 for rpm. A 60-pulse-per-revolution sensor reading 1,800 Hz corresponds to 30 Hz mechanical frequency, or 1,800 rpm. Confirm effective counts per revolution when using quadrature decoding.
  3. Check the signals together. Put vibration and speed on the same synchronised acquisition time base. For the heatmap, make both available in the same live source stream. At a known steady speed, confirm that the speed channel is correct and that a known rotation-related component appears where expected in an FFT. Record the raw vibration and speed if you will need to analyse the run again.

See Measuring with a quadrature encoder and Frequency measurement with the Chronos method for input configuration. Confirm the encoder electrical interface against the selected module.

2. Calculate order values on a Q.station

Use this route when you need selected order results as controller variables for logging or trending. It is separate from configuring the dashboard heatmap. Confirm that the installed controller firmware and GI.bench build expose the Order Tracking evaluator.

  1. Open the Q.station Controller Settings and configure the FFT Processor for the vibration channel. Select the FFT size, overlap and window. Set up writable result variables before assigning them to evaluators. See How to perform FFT and spectral analysis for the base procedure.
  2. In the processor’s order tracking settings, use Based on mechanical frequency when you have a measured speed reference. Assign the Frequency input in mechanical Hz, not rpm or raw pulse frequency. The processor searches near the expected frequency of each order. The alternative Based on maxima finding with method Gauss detects peaks with smoothing and a threshold; without a shaft reference, verify that the selected peaks really represent the shaft orders of interest.
  3. Set Evaluator count to 1 to start and select Order tracking as the evaluator type. In this mode, Start frequency is the order number, despite the frequency label. Enter 1 for 1×. Stop frequency is not used for this evaluator. Assign the first writable result variable to amplitude and the second to the detected frequency in Hz. Add further evaluators only after verifying 1×.
  4. Set Range for calculation narrowly enough to exclude unrelated peaks. The demonstration used 2 for a search of ±2 FFT bins around the expected frequency. It also used a Hanning window, 0% buffer overlap, 0 integrations and Magnitude representation. These are example settings, not defaults. With zero integrations the result remains in the acceleration domain when acceleration is the source. Choose Magnitude or RMS consistently and label your results.
  5. Apply the configuration to the controller. Verify the result at steady speed and during a controlled sweep. Add the numeric results to a Yt trend or to an independent logger if required. This records the values you assign; it does not automatically archive the dashboard heatmap.

Choosing an FFT size: For N samples at sample rate fs, the block lasts N/fs and the FFT bin spacing is fs/N. Longer blocks separate nearby peaks better but can smear orders if speed changes substantially during the block. Shorter blocks reduce that speed change at the expense of frequency resolution. Overlap gives more frequent updates but does not shorten the block. Start with one channel and validate the update rate before adding more processors.

3. Build an order tracking dashboard in GI.bench

  1. Open the dashboard in edit mode and add an Order Tracking Heatmap. Drag in the raw vibration channel first as Value and the shaft-speed channel second as Trigger. Do not use the controller’s scalar 1× result as the heatmap’s vibration input.
  2. In the heatmap’s Trigger settings, set Start Value, Stop Value and Increment in the units of the trigger channel. Select Between for a rising sweep between two limits or Below Limits for a falling sweep through two limits. Apply and save. For a run-up example with a trigger in rpm, use 1,200 to 3,600 rpm and a 100 rpm increment; start below 1,200 rpm so the trigger crosses the start value. These are example settings, not defaults.
  3. Keep the sweep slow enough to resolve the peaks and capture the required speed steps. In Element Settings, Points is the FFT length and Window sets the weighting. The demonstration used 4,096 points and Hanning. Read the map as frequency in Hz on the horizontal axis, the trigger value on the vertical axis and vibration amplitude in colour.
  4. In Peak Finder, enable Use trigger as mechanical frequency when the trigger is the shaft frequency in Hz. If the available channel is in rpm, create a scaled channel (rpm ÷ 60) and enter the limits and increment in Hz for that channel. With the option off, the peak finder uses Gaussian smoothing and a peak threshold; verify any selected peaks against the measured speed.
  5. Add a Yt chart for shaft speed and raw vibration. This lets you check the direction of the sweep, clipping and interruptions. For an order cut (amplitude of one order versus speed), configure the desired FFT Evaluators in the first, vibration variable’s dashboard settings and link an XY Chart to the heatmap. The linked cut comes from the heatmap data. If you need a separate plot of the controller’s order result, plot that result against speed as an ordinary XY data pair.

How to read the plot: With Hz on the horizontal axis and rpm on the vertical axis, orders appear as diagonal ridges, while a fixed-frequency component stays near one horizontal coordinate. For example, a 2× ridge crosses 100 Hz at 3,000 rpm. If its amplitude grows there, compare neighbouring orders and repeat the sweep before attributing the feature to a resonance.

4. Monitor an order against a live alarm limit

Use the controller’s order-amplitude result as the input to your alarm logic. Compare it with a threshold in the same unit and with the same peak or RMS convention. Display the threshold and alarm state alongside the order trend and shaft speed. In the Q.screen demonstration, the test.con display also showed broadband vibration next to the selected order amplitude, so an operator could distinguish general vibration from a rise in one order.

Verify the alarm above and below the setpoint with a known signal and check the response time on the configured system. The result updates as FFT blocks are processed, so the FFT block length and overlap affect the response. Use this for monitoring; do not substitute it for a safety-rated protection system.

Verify the measurement before relying on it

  • No spectra appear: Check the Value and Trigger assignments, confirm both are live and from the same stream, and check the trigger mode and crossing direction. In Between mode a new run-up must first fall below the start value and then cross it again from below.
  • The 1× trace is in the wrong place: Check pulses per revolution, any gearbox ratio, the shaft chosen as the reference, and Hz versus rpm scaling.
  • Order peaks broaden or disappear: Check for speed dropouts, clipping and inadequate bandwidth. Compare a slower sweep or shorter FFT block, then review the search range and smoothing. Verify amplitude convention before comparing values across charts.
  • Results stop updating with several processors: Check processor health and buffer overruns, then repeat with one processor and compare. Validate the final channel count and FFT settings on the deployed controller and firmware.

Frequently asked questions about order tracking

Do I need an encoder?

No. A suitable tachometer or other synchronised speed signal can provide the reference. An encoder may be useful when you need better speed resolution or a separate angle measurement. Peak finding without a shaft reference is possible, but does not by itself identify shaft orders.

Does this workflow resample the signal at fixed shaft angles?

The workflow described here uses fixed-rate, time-domain data and FFT spectra. Use a suitable angle-synchronous analysis method if your test specifically requires angle resampling or high-fidelity order separation during rapid speed changes.

What should I record for later analysis?

Keep the raw vibration and shaft reference together, with their sampling rates, units and the reference shaft identified. Log any controller evaluator results you need separately. Check how your software version handles historical visualisation rather than assuming the live heatmap is saved with the measurement.